PCAN MU-Thermocouple1 CAN
SKU #: IPEH-002205-J, IPEH-002205-K, IPEH-002205-T
Configurable System for Data Acquisition and Data Processing
The Thermocouple Measuring Unit offers connections for 8 thermo-elements for different temperature ranges (J, K, T).
Measurement data can be pre-processed and transmitted via CAN bus using a central microcontroller. Configuration work involves using Windows® software on a computer connected to the same CAN bus.
Scope of supply
- MU-Thermocouple1 CAN in aluminum casing
- Mating connector for voltage supply
- Configuration software Thermocouple Configuration for Windows® 10, 8, 7 (32/64-bit) (Configuration requires a PEAK CAN interface
- Configuration software PPCAN-Editor 2 for Windows® Details ...
- Manual in PDF format
- 8 Mini sockets for thermocouple types J, K, and T (depending on the assembly of measuring modules at delivery)
- 4 galvanically isolated measuring modules, each with 2 thermocouple sockets of the same type
- Measuring ranges:
- J: -210 to +1121 °C
- K: -200 to +1370 °C
- T: -200 to +400 °C
- Measurement accuracy: 0.2 %
- Accuracy of the reference sensors: typically ±0.5 K, max. ±1.0 K at +25 °C ambient temperature
- Resolution for temperature data at CAN communication: 1/16 °C
- High-speed CAN connection (ISO 11898-2) for data transfer and configuring
- Bit rates from 40 kbit/s up to 1 Mbit/s
- Galvanic isolation up to 500 V
- NXP TJA1040 CAN transceiver
- Complies with CAN specifications 2.0 A/B
- Basic configuration with the Windows® software Thermocouple Configuration
- Advanced configuration with the Windows® software PPCAN-Editor 2
- Pre-processing of readings possible with integrated microcontroller
- Aluminum casing with flange. DIN rail fixing option available on request
- Voltage supply from 6 to 34 V
- Extended operating temperature range from -40 to 85 °C (-40 to 185 °F)
Power Supply | Supply voltage: Nominal 12 V DC (6 - 34 V possible) Current consumption: ca. 100 mA at 12 V |
---|---|
Measuring modules | Count: 4 Connectors: 2 Mini sockets for thermocouples (DIN EN 50212) per module, 8 altogether Galvanic isolation: each measuring module internally against the supply voltage, up to 500 V Supported thermocouple types (IEC 60584-1): J (-210 - +1121 °C); K (-200 - +1370 °C); T (-200 - +400 °C); (according to the used module) Measurement accuracy for thermocouple inputs: ±0.2 % Influence of ambient temperature: 10.5 ppm/K Reference sensors: 4 (1 per measuring module) Accuracy of the reference sensors: ±1 K at 0 - +70 °C; ±2 K at -20 - +85 °C; ±3 K at -40 - +125 °C Measuring category: CAT I (only electric circuits that are not connected to the mains) |
Controller module | Microcontroller: NXP LPC2366 Resolution for temperature data at CAN communication: 1/16 °C Sampling rate of the sensors: 3 Hz |
CAN | Specification: ISO 11898-2, High-speed CAN 2.0A (standard format) and 2.0B (extended format) Bit rates: 40 kbit/s - 1 Mbit/s; Lower bit rates on request Transceiver: NXP TJA1040T Connection: D-Sub (m), 9-pin, assignment according to specification CiA® 102; Optional 5-Volt supply at pin 1 for external devices (e.g. bus converters), max. 100 mA; Galvanic isolation up to 500 V Termination: none |
Dimensions | 130 x 60 x 73 mm (W x H x D) (incl. mounting plate and connectors) See also dimension drawings in Appendix B in the User's Manual |
Weight | 420 g |
Operating temperature | -40 - +85 °C (-40 - +185 °F) |
Storage temperature | -40 - +100 °C (-40 - +212 °F) |
Relative humidity | 15% to 90% (non-condensing) |
Ingress protection (IEC 60529) | IP20 |
EMC | DIN EN 61326-1; EC directive 2004/108/EG |
Safety | EN 61010-1 + Amendments 1 and 2 |
MU-Thermocouple1 CAN manual |
User's Guide in PDF format |
PPCAN-Editor Configuration Tutorial For MU-Thermocouple1 CAN |
PPCAN-Editor configuration tutorial for MU-Thermocouple1 CAN Manual |
PPCAN-Editor Configuration Tutorial Example Files |
Configuration Tutorial Example Files |
PPCAN-Editor Function Block Reference Guide |
PPCAN-Editor function block reference guide in PDF format |
Thermocouple Configuration |
Configuration software for MU-Thermocouple1 CAN for Windows® 11 (64-bit), 10 (32/64-bit). |
PPCAN-Editor 2 |
Configuration software for PCAN-MIO and PCAN-Router Pro for Windows® 11 (64-bit), 10 (32/64-bit). |
PEAK-Flash |
Easy-to-use tool for updating the firmware of PEAK-System hardware products as well as CAN and LIN adapters for Windows® 11 (64-bit), 10 (32/64-bit). |
Configurable System for Data Acquisition and Data Processing
The Thermocouple Measuring Unit offers connections for 8 thermo-elements for different temperature ranges (J, K, T).Measurement data can...